GLIDESHEATH SLENDER A-KIT 6F 10CM .021GW. 5 units per box. Product Code: 60-1060. The Glidesheath Slender is used to facilitate placing a catheter through the skin into the radial artery. Recall
GLIDESHEATH SLENDER A-KIT 6F 10CM .021GW. 5 units per box. Product Code: 60-1060. The Glidesheath Slender is used to facilitate placing a catheter through the skin into the radial artery.
FDA recall Class II
Ongoing
FDA lists this Class II recall for GLIDESHEATH SLENDER A-KIT 6F 10CM .021GW. 5 units per box. Product Code: 60-1060. The Glidesheath Slender is used to facilitate placing a catheter through the skin into the radial artery.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2930-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 15, 2026
- FDA report date
- Aug 19, 2026
- Quantity in commerce
- 7,490 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
GLIDESHEATH SLENDER A-KIT 6F 10CM .021GW. 5 units per box. Product Code: 60-1060. The Glidesheath Slender is used to facilitate placing a catheter through the skin into the radial artery.
Affected model & UDI checker
Search the exact identifiers published for Z-2930-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | GLIDESHEATH SLENDER A-KIT 6F 10CM .021GW. 5 units per box. Product Code: 60-1060. The Glidesheath Slender is used to facilitate placing a catheter through the skin into the radial artery. | Check source notice | 30389701011975 |
| Not specified | GLIDESHEATH SLENDER A-KIT 6F 10CM .021GW. 5 units per box. Product Code: 60-1060. The Glidesheath Slender is used to facilitate placing a catheter through the skin into the radial artery. | Check source notice | 00389701011974 |
FDA code information
Product Code: 60-1060. Shelfbox UDI-DI: 30389701011975; Pouch UDI-DI: 00389701011974; UDI-PI: (17)280501(10)0001441871. Lot Number: 0001441871.
Why was this product recalled?
Product was packaged with the 0.025 guidewire instead of the 0.021 guidewire.
What actions did Terumo Medical Corporation take?
Terumo Medical Corporation notified consignees on about 06/15/2026. Consignee were instructed to immediately isolate the affected lot number, 0001441871, arrange for the return of those affected units, notify any facilities or customers to which affected units have been further transferred or distributed, and complete and return the provided response form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2930-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026