Straumann Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim (Article 021.5314); implant, endosseous, root-form Recall
Straumann Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim (Article 021.5314); implant, endosseous, root-form
FDA recall Class II
Ongoing
FDA lists this Class II recall for Straumann Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim (Article 021.5314); implant, endosseous, root-form. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2935-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 16, 2026
- FDA report date
- Aug 19, 2026
- Quantity in commerce
- 175 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Straumann Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim (Article 021.5314); implant, endosseous, root-form
Affected model & UDI checker
Search the exact identifiers published for Z-2935-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Straumann Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim (Article 021.5314); implant, endosseous, root-form | Check source notice | 7630031707055 |
FDA code information
UDI/DI 7630031707055, Lot ZVFV5
Why was this product recalled?
Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim may inadvertently contain implants that are 12mm in length instead of 14mm as the labeling would suggest.
What actions did Straumann USA LLC take?
Straumann Group issued an Urgent Field Safety Action notice to its consignees on 7/16/2026 via UPS. The notice explained the issue, potential risk, and requested the following: "Action to be taken: 1. Identify and segregate any units in your stock with carton packaging labelled as Article 021.5314 Bone Level Tapered Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim" (Article 021.5314, Lot ZVFV5) 2. Return any segregated units with carton packaging labelled as Article 021.5314 Bone Level Tapered Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim", Lot ZVFV5 for replacement or refund 3. If the product has been implanted with no unexpected clinical occurrence during implantation, then no special follow up outside of the dentist s standard follow procedure is necessary 4. In all cases, complete and return the enclosed Customer Confirmation Form with the affected product using the return UPS label provided to: Straumann USA, LLC 60 Minuteman Road Andover, MA 01810 If product is not being returned, the completed Customer Confirmation Form can be scanned and emailed to [email protected] Transmission of the Field Safety Notice This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the affected devices have been transferred."
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2935-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.