FDA RECALL RECORD · Z-2936-2026

BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based multiplexed nucleic acid test, in vitro diagnostic Recall

BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based multiplexed nucleic acid test, in vitro diagnostic

FDA recall Class II

Ongoing

FDA lists this Class II recall for BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based multiplexed nucleic acid test, in vitro diagnostic. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2936-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 24, 2026
FDA report date
Aug 19, 2026
Quantity in commerce
2,370 pouches (79 kits)
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based multiplexed nucleic acid test, in vitro diagnostic

Affected model & UDI checker

Search the exact identifiers published for Z-2936-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
423742BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based multiplexed nucleic acid test, in vitro diagnosticCheck source noticeNot specified
Not specifiedBIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based multiplexed nucleic acid test, in vitro diagnosticCheck source notice00815381020529
FDA code information

Part Number: 423742 UDI-DI code: 00815381020529 Pouch Lot: 3VLB25 Kit Lot: 1872125

View the original FDA recall notice

Why was this product recalled?

Due to increase in false negative results

What actions did Biofire Diagnostics, LLC take?

On 06/24/2026, an "URGENT MEDICAL DEVICE CORRECTION" Letter was emailed to customers informing them that bioMrieux received complaints of false-negative results associated with BIOFIRE RP2.1 kit lot 1872125 / pouch lot 3VLB25. An internal investigation confirmed the presence of incorrect PCR1 primers in a subset of pouches within BIOFIRE RP2.1 kit lot 1872125 introduced during reagent manufacturing. This field safety notice was issued because the incorrect PCR1 primers will lead to pouch pathogen assay negative results with passing control assays. Customers are instructed to: "Examine your inventory for the lot identified in this field safety notice. "Cease use and discard any remaining product from this lot in your possession. "Confirm the quantity of kits that have been destroyed, previously used, or returned on the attached Acknowledgement of Receipt Form. "If you have further distributed this product, please identify any recipients and notify them at once. "Complete the accompanying Acknowledgement of Receipt Form and return to bioMrieux so that bioMrieux may acknowledge your receipt of this notification. "bioMerieux will be providing pouch credit for affected products. For Questions - contact local Customer Service Representative at (800) 682-2666. For non-urgent inquiries, contact the Technical Support Center by email at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology