ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System. Recall

ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.

FDA recall Class II

Ongoing

FDA lists this Class II recall for ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2937-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 18, 2026
FDA report date
Aug 19, 2026
Quantity in commerce
400 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.

Affected model & UDI checker

Search the exact identifiers published for Z-2937-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
ST101UBProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.Check source noticeNot specified
Not specifiedProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.Check source notice5711435000156
FDA code information

Model number: ST101UB / 08787174 190; GTIN code: 5711435000156; Lot-number: 04909125 and 04912125.

View the original FDA recall notice

Why was this product recalled?

The use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in intended users.

What actions did Bioporto Diagnostics A/S take?

A Quality Notification dated 6/3/26 was sent to consignees via email notifying them of the recall. Consignees in the US are to coordinate replacement of devices. ROW Consignees are to scrap unopened kits in accordance with local requirements and request replacements.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology