ERBECRYO 2 Cryosurgical Unit Recall
ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
FDA recall Class II
Ongoing
FDA lists this Class II recall for ERBECRYO 2 Cryosurgical Unit. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2938-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 26, 2026
- FDA report date
- Aug 19, 2026
- Quantity in commerce
- 1126 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
Affected model & UDI checker
Search the exact identifiers published for Z-2938-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 10402-000 | ERBECRYO 2 Cryosurgical Unit | Check source notice | Not specified |
| Not specified | ERBECRYO 2 Cryosurgical Unit | Check source notice | 04050147013797 |
FDA code information
UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3
Why was this product recalled?
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
What actions did Erbe USA Inc take?
On June 26, 2026, Erbe USA issued a Urgent: Medical Device Correction update notification to affected consignees via multiple methods ( Email/Website/Hand delivery). ERBE ask consignees to take the following actions: 1. We request the ERBECRYO 2 Unit be made available to the Erbe Representative to install the (V)1.0.4 Software. The Erbe Representative will perform all tasks required to update the software. 2. Confirm acknowledgement with Erbe Representative on-site. 3. Share this notice with all users within your facility. 4.While awaiting software update, there is no change of the Cryosurgical Unit s intended use or performance. Please continue to follow all warnings, instructions, guidelines, etc. involved with the Unit and Cryoprobe.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2938-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.