Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2954-2026

Dentsply Sirona Capsule Mixer Model No. 5546068 Recall

Dentsply Sirona Capsule Mixer Model No. 5546068

FDA recall Class II

Ongoing

FDA lists this Class II recall for Dentsply Sirona Capsule Mixer Model No. 5546068. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2954-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 20, 2026
FDA report date
Aug 26, 2026
Quantity in commerce
1,444 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Dentsply Sirona Capsule Mixer Model No. 5546068

Affected model & UDI checker

Search the exact identifiers published for Z-2954-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
5546068Dentsply Sirona Capsule Mixer Model No. 5546068Check source noticeNot specified
Not specifiedDentsply Sirona Capsule Mixer Model No. 5546068Check source notice9336472005542
FDA code information

UDI:9336472005542; Serial Numbers 0003-1533

View the original FDA recall notice

Why was this product recalled?

Potential for electrical discharge from unit when unplugged.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did SDI Limited take?

On July 20, 2026, Dentsply Sirona issued a Urgent Medical Device Recall notice to affected consignees via mail. Dentsply Sirona ask consignees to take the following actions: Users 1. Switch off the unit using the rear power switch before removing the plug from the main socket 2. Avoid immediate contact with the plug pins after unplugging. 3. The users should return their unit to Dentsply Sirona to have a corrective modification installed. Please contact Racheal Elliott at [email protected]. 4. Please ensure that all relevant personnel within your organization are informed of the notice. If you have supplied affected units to another organization, please provide this with a copy of the notification. Dealers: 1.Identify all customers who have received the affected units and immediate forward the accompanying communication to those customers. 2.Ensure that all relevant personnel within your organization are informed of this notice. 3.Complete the dealer reply form provided on pages 2 and 3 and rturn it to Dentsply Sirona. 4. Return the completed form via email to Racheal Elliott at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.