FDA RECALL RECORD · Z-2956-2026

Cardiac Workstation 7000. Product Number: 860441. Recall

Cardiac Workstation 7000. Product Number: 860441.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Cardiac Workstation 7000. Product Number: 860441.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2956-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 1, 2026
FDA report date
Aug 26, 2026
Quantity in commerce
1,810 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Cardiac Workstation 7000. Product Number: 860441.

Affected model & UDI checker

Search the exact identifiers published for Z-2956-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedCardiac Workstation 7000. Product Number: 860441.Check source notice00884838094833
FDA code information

Product Number: 860441. UDI: (01)00884838094833. Serial Numbers: US22520232, US32520244, US32520248, US42520323, US42520320, US42520300, US52520333, US52520332, US52520345, US52520382, US52520359, US52520390, US52520373, US52520412, US52520420, US52520426, US52520461, US62520469, US62520470, US62520474, US62520476, US62520483, US72520504, US72520518, US72520531, US62520499, US72520500, US72520648, US72520526, US72520524, US72520522, US72520517, US72520590, US72520547, US72520592, US82520720, US72520603, US72520640, US72520580, US72520666, US82520755, US82520752, US82520762, US72520662, US72520652, US72520664, US82520771, US82520844, US82520782, US82520774, US82520769, US82520777, US82520775, US82520855, US82520820, US82520823, US82520824, US82520744, US72520678, US82520822, US82520821, US82520726, US72520667, US72520670, US72520668, US82520848, US72520669, US92520885, US82520857, US82520851, US82520863, US92520881, US92520890, US92520901, US92520903, US92520894, US92520905, US92520906, US92520981, USO2521043, USO2521042, USO2521050, USO2521128, USO2521092, USO2521122, USO2521065, USN2521141, US92520990, USN2521132, USN2521144, USN2521149, USN2521130, USN2521179, USN2521150, USN2521161, USN2521163, USN2521164, USN2521160, USO2521072, USN2521175, USN2521170, USN2521176, USN2521185, USN2521206, USN2521211, USN2521214, USN2521213, USN2521196, USN2521226, USN2521223, USN2521249, USN2521215, USN2521224, USN2521246, USN2521243, USN2521235, USN2521221, USN2521222, USN2521342, USN2521198, USN2521254, USN2521252, USN2521294, USN2521286, USN2521258, USN2521257, USN2521256, USN2521307, USN2521288, USN2521287, USD2521394, USN2521289, USN2521300, USN2521296, USD2521365, USN2521302, USD2521427, USD2521407, USD2521436, USD2521433, USD2521448, USD2521440, USD2521475, USD2521430, USN2521308, USD2521362, USD2521374, USN2521309, USD2521355, US12621548, US12621550, US12621553, US12621671, US12621672, US12621629, US12621713, US12621712, US12621711, US12621710, US22621762, US22621763, US22621777, US22621787, US22621796, US22621859, US32621881, US32621887, US32621893, US32621905, US32621886, US32621884, US32621894, US32621906, US32621907, US32621909, US32621910, USD2521420, USD2521409, USD2521411, USD2521416, USD2521422, US32621911, US32621918, US32621912, US32621913, US32621936, US32622017, US32621923, US32621980, US32621953, US32621952, US32621950, US32621954, US32621895, US32622011, US32622018, US32622051, US32622023, US32622077, US32622078, US32622046, US42622153, US32622027, US32622053, US32622052, US32622076, US32622064, US42622108, US32622065, US42622129, US42622123, US42622110, US42622148, US42622113, US42622111, US42622154, US42622147, US42622139, US42622146, US42622116, US42622149, US42622150, US42622222, US42622188, US42622183, US42622187, US42622156, US42622184, CN72420050, CN82420071, CNO2420085, CNN2420093, CNN2420096, CNN2420111, CNN2420115, CND2420119, CND2420145, CND2420134, CND2420150, CND2420151, CND2420168, CND2420167, CND2420169, CND2420171, CN12520174, CN22520206, CN22520215, CN22520218, CN22520214, CN22520226, CN22520242, CN32520246, CN42520274, CN42520317, CN42520291, CN52520389, CN52520379, CN52520445, CN52520460, CN52520454, CN62520466, CN62520487, CN72520565, CN72520650, CN72520631, CN82520721, CN72520651, CN82520758, CN82520764, CN82520791, CN82520792, CN82520793, CN82520837, CN82520842, CN82520864, CN82520809, CN92520880, CN92520871, CN92520923, CN92520929, CN92520970, CN92520935, CN92520979, CN92520976, CNO2521015, CNO2520997, CNO2521008, CNO2521007, CNO2521026, CNO2521030, CNO2520998, CNN2521156, CNN2521212, USN2521219, CNN2521230, CNN2521283, CNN2521260, USN2521291, CNN2521346, CNN2521345, CNN2521344, CNN2521343, CND2521375, CND2521352, CND2521415, CND2521439, USD2521429, CND2521506, CN12621557, CND2521535, CND2521544, CN12621562, CN12621580, CN12621577, CN12621576, CN12621579, CN12621690, CN12621687, CN12621696, CN12621697, CN12621698, CN22621734, CN22621799, CN22621798, CN32621877, CN22621874, CN32621875, CN32621888, CN32621890, CN32621896, CN32621968, CN32621997, CN32622002, CN32621990, CN32621983, CN32622007, CN32622008, CN32622005, CN32622006, CN32622021, CN32622047, CN32622066, CN32622075, CN42622107, CN42622124, CN42622134, CN42622220, CN42622221

View the original FDA recall notice

Why was this product recalled?

Device will intermittently provide inaccurate data related to native heart rate.

What actions did Philips North America Llc take?

Philips notified consignees on about 07/01/2026 via URGENT Medical Device Correction letter. Consignees were informed that they could continue the use of the device but to refer to the information within the notification letter, disable the automatic Standby feature on the Cardiac Workstation and avoid manually putting the Cardiac Workstation in Standby mode to minimize the inaccurate heart rate value from occurring, and to pass the notice to all those who need to be aware within each organization or to any organization where affected product(s) have been potentially transferred. Additionally, consignees were requested to place the Urgent Medical Device Correction letter with the documentation of the Cardiac Workstation and associated devices in a place where it is most likely to be seen and viewed. Philips representatives will contact consignees to arrange an upgrade to the Cardiac Workstation Release 1.2.0.7.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.