Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2965-2026

Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D)/10471981 IVD, RxOnly Recall

Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D)/10471981 IVD, RxOnly

FDA recall Class II

Ongoing

FDA lists this Class II recall for Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D)/10471981 IVD, RxOnly. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2965-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 13, 2026
FDA report date
Aug 26, 2026
Quantity in commerce
24 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D)/10471981 IVD, RxOnly

Affected model & UDI checker

Search the exact identifiers published for Z-2965-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedSiemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D)/10471981 IVD, RxOnlyCheck source notice00630414976891
FDA code information

Lot Code: D387/UDI: 00630414976891

View the original FDA recall notice

Why was this product recalled?

Syphilis Screen assay may contain a negative patient bias resulting in low reactive patient results being reported as nonreactive.

What actions did Siemens Healthcare Diagnostics Products, Ltd. take?

On July 13, 2026, Siemens Healthineers issued a Urgent Medical Device Correction to affected consignees via FedEx. Siemens asked consignees to take the following actions: 1. Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2.Discontinue use and discard the affected kit lot. 3. Review your current inventory to determine your laboratory s replacement needs and provide this information to Siemens Healthineers for reporting to the authorities. 4. Complete and return the Field Correction Effectiveness Check Form attached to this letter within (30) days. 5.Retain this letter with your laboratory records and forward this letter to those who may have received the affected product(s).

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology