Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2970-2026

Blanketrol III, Model 233; Thermal regulating system. Recall

Blanketrol III, Model 233; Thermal regulating system.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Blanketrol III, Model 233; Thermal regulating system.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2970-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 13, 2026
FDA report date
Aug 26, 2026
Quantity in commerce
9714 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Blanketrol III, Model 233; Thermal regulating system.

Affected model & UDI checker

Search the exact identifiers published for Z-2970-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
233Blanketrol III, Model 233; Thermal regulating system.Check source noticeNot specified
Not specifiedBlanketrol III, Model 233; Thermal regulating system.Check source notice1061303186102
Not specifiedBlanketrol III, Model 233; Thermal regulating system.Check source notice1061303186107
Not specifiedBlanketrol III, Model 233; Thermal regulating system.Check source notice1061303186139
Not specifiedBlanketrol III, Model 233; Thermal regulating system.Check source notice1061303187107
Not specifiedBlanketrol III, Model 233; Thermal regulating system.Check source notice1061303187307
FDA code information

UDI: 1061303186102, UDI: 1061303186107, UDI: 1061303186139, UDI: 1061303187107, UDI: 1061303187307, Serial Numbers:131-2-00580 through 262-3-20277

View the original FDA recall notice

Why was this product recalled?

Overheating of the 115V CPU board may cause the device to shut down.

What actions did Gentherm Medical, LLC take?

Genthem issued an URGENT Medical Device Notice to its consignees on 7/17/2026 via email. The notice explained the issue, potential risk, and request the following: "Interim Recommendations: 1. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available. 2. Users should review and follow the device Instructions for Use; users should routinely verify the device s function and patient s condition during use. 3. Ensure that all users are informed of the contents of this letter. If you have further distributed this product, please provide those accounts with a copy of this notice." For questions or if you have experienced the described issue or a related adverse event, call Gentherm Medical Technical Support at 1-888-437-5608.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.