Custom Made Device Implant 6123 Navicular Cage Small Right Rev A, Model/Catalog Number 6123S R; foot implant Recall
Custom Made Device Implant 6123 Navicular Cage Small Right Rev A, Model/Catalog Number 6123S R; foot implant
FDA recall Class III
Ongoing
FDA lists this Class III recall for Custom Made Device Implant 6123 Navicular Cage Small Right Rev A, Model/Catalog Number 6123S R; foot implant. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2971-2026
- Recall classification
- Class III
- Enforcement status
- Ongoing
- Initiation date
- Jul 10, 2026
- FDA report date
- Aug 26, 2026
- Quantity in commerce
- 1
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Custom Made Device Implant 6123 Navicular Cage Small Right Rev A, Model/Catalog Number 6123S R; foot implant
Affected model & UDI checker
Search the exact identifiers published for Z-2971-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 6123S | Custom Made Device Implant 6123 Navicular Cage Small Right Rev A, Model/Catalog Number 6123S R; foot implant | Check source notice | Not specified |
FDA code information
Lot Number AOCIA6123
Why was this product recalled?
Device was recalled due to an elevated oxygen level in excess of the allowable limit. Elevated oxygen levels may contribute to reduced material ductility.
What actions did Paragon 28, Inc. take?
Paragon 28 issued an URGENT MEDICAL DEVICE RECALL to its consignees on 7/14/2026 via FedEx 2nd day. The notice explained the issue, potential risk, and requested the following: Distributors: "Your Responsibilities 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form for each return and send to [email protected]. This form must be returned even if you do not have affected products available to return in your territory. b. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. c. Mark RECALL on the outside of the returned cartons. 4. Retain a copy of your Inventory Return Certification and product return forms for your records in the event of a compliance audit of your facility. 5. If you have further questions or concerns after reviewing this notice, please call 888-728-1888. Alternatively, your questions may be emailed to [email protected]." Risk Manager: "Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Paragon 28 sales representative and quarantine all affected product. Your Paragon 28 sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retai
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2971-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026