Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL Recall
Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
FDA recall Class II
Ongoing
FDA lists this Class II recall for Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2976-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 13, 2026
- FDA report date
- Aug 26, 2026
- Quantity in commerce
- 55 units
- Distribution
- See key details
- FDA cause
- Unknown/Undetermined by firm
What is being recalled?
Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
Affected model & UDI checker
Search the exact identifiers published for Z-2976-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| OSR4Q229 | Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL | Check source notice | Not specified |
| Not specified | Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL | Check source notice | 00842768023293 |
FDA code information
Lots: 2597, 2627, 2709/UDI: 00842768023293
Why was this product recalled?
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Siemens Healthcare Diagnostics, Inc. take?
On July 14, 2026, Simens Healthineers issued a Urgent Medical Device Correction notification via FedEx. Simens asked consignees to take the following actions: 1. Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2. Discontinue use of and discard the kit lots listed in the Products Section. 3. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. 4. Retain this letter with your laboratory records and forward this letter to those who may have received this product.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2976-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology