Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2981-2026

MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782; Recall

MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;

FDA recall Class II

Ongoing

FDA lists this Class II recall for MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2981-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 13, 2026
FDA report date
Aug 26, 2026
Quantity in commerce
7
Distribution
See key details
FDA cause
Process design

What is being recalled?

MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;

Affected model & UDI checker

Search the exact identifiers published for Z-2981-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MMT-1781MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;Check source noticeNot specified
MMT-1782MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;Check source noticeNot specified
FDA code information

Model/UDI-DI Serial Numbers: MMT-1781/00763000404536/NG2665223H, NG2748200H, NG2748664H, NG2755237H, NG2716176H, NG3547582H. MMT-1782/00763000404543/NG2768190H

View the original FDA recall notice

Why was this product recalled?

Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medtronic MiniMed, Inc. take?

On 7/13/2026, field safety notices were sent to customers asking them to do the following: 1) Stop using the loaner pump and contact the firm to arrange for loaner pump return. 2) Contact your Healthcare Professional (HCP) to ensure you have access to backup insulin therapy supplies or alternative plans, which may include short- and long-acting insulins, syringes, with dosage instructions from your HCP. 3) Complete and return the customer acknowledgement form. Questions can be directed to the firm's Tech Support: 1-800-646-4633

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.