Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131 Recall

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

FDA recall Class II

Ongoing

FDA lists this Class II recall for Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2983-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 8, 2026
FDA report date
Aug 26, 2026
Quantity in commerce
Not reported
Distribution
See key details
FDA cause
Component design/selection

What is being recalled?

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Affected model & UDI checker

Search the exact identifiers published for Z-2983-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
P2131Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131Check source noticeNot specified
FDA code information

Lot Code: All serial numbers/ GTIN: (01) 00862992000327

View the original FDA recall notice

Why was this product recalled?

The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Fort Defiance Industries, LLC take?

Between the dates of July 8, 2026 - July 16, 2026 Fort Defiance Industries (FDI) issued a Urgent Medical Device Recall to affected consignees vi Email. FDI ask consignees to take the following actions: 1. Devices can continue in operation until the fix is implemented. 2. In the interim, users should avoid touching the WRS power connector during operation as a precaution. 3. Please contact FDI and order the upgrade kit complete with instructions to perform this Field Correction. 4. After receiving the kit, please review the instructions and determine if you are competent to perform the update. The instructions will state alternate options to complete the repair (i.e. send to military depot). 5. Perform the repair 6. FDI requests a phone call or e-mail confirmation when the upgrade is completed. This confirmation can be a phone call or e-mail message.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.