Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131 Recall
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
FDA recall Class II
Ongoing
FDA lists this Class II recall for Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2983-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 8, 2026
- FDA report date
- Aug 26, 2026
- Quantity in commerce
- Not reported
- Distribution
- See key details
- FDA cause
- Component design/selection
What is being recalled?
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Affected model & UDI checker
Search the exact identifiers published for Z-2983-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| P2131 | Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131 | Check source notice | Not specified |
FDA code information
Lot Code: All serial numbers/ GTIN: (01) 00862992000327
Why was this product recalled?
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Fort Defiance Industries, LLC take?
Between the dates of July 8, 2026 - July 16, 2026 Fort Defiance Industries (FDI) issued a Urgent Medical Device Recall to affected consignees vi Email. FDI ask consignees to take the following actions: 1. Devices can continue in operation until the fix is implemented. 2. In the interim, users should avoid touching the WRS power connector during operation as a precaution. 3. Please contact FDI and order the upgrade kit complete with instructions to perform this Field Correction. 4. After receiving the kit, please review the instructions and determine if you are competent to perform the update. The instructions will state alternate options to complete the repair (i.e. send to military depot). 5. Perform the repair 6. FDI requests a phone call or e-mail confirmation when the upgrade is completed. This confirmation can be a phone call or e-mail message.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2983-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.