restor3d Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Recall

Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2987-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 30, 2026
FDA report date
Aug 26, 2026
Quantity in commerce
59
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder

Affected model & UDI checker

Search the exact identifiers published for Z-2987-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1500-1030Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid BaseplateCheck source noticeNot specified
1500-1035Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid BaseplateCheck source noticeNot specified
Not specifiedBrand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid BaseplateCheck source notice00840097513225
Not specifiedBrand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid BaseplateCheck source notice00840097513232
FDA code information

Lot Code: Model No 1500-1030; UDI-DI 00840097513225; Lot Numbers A00030, A00065, A09828, A09830 Model No 1500-1035; UDI-DI 00840097513232; Lot Numbers A00064, A09829

View the original FDA recall notice

Why was this product recalled?

Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did restor3d Inc. take?

On 06/30/2027 and 07/02/2026, emailed an "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" Letter informing customers that, restor3d has identified a machining nonconformance affecting the peripheral screw holes on certain Veritas standard glenoid baseplates. Customers are instructed to: Immediately upon receipt of this notice: .Do not implant or otherwise use any affected units identified above. .Quarantine all affected, unused units identified above and remove them from implant sets, consignment inventory, inventory shelves, and any other storage locations. . Complete the attached Field Action Response Form (Attachment A) and return it to restor3d within 10 business days, whether or not affected product is in your possession, so that restor3d can confirm effective distribution of this notice. . Package the affected Veritas standard glenoid baseplates (1500-1030 and 1500-1035) to return to restor3d. restor3d has provided a prepaid FedEx shipping label by email for shipment back to 4001 NC-54 Hwy, Suite 2160 Durham, NC 27709. . Promptly report to restor3d any intraoperative issue, patient adverse event, suspected device related complication, loosening, migration, revision surgery, or other clinical concern potentially associated with the affected product.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

Product / regulatory recordCompanyPathwayDecision dateSource
Restor3D510(k) clearanceMay 19, 2025FDA

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.