Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2991-2026

Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit Recall

Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit. Respiratory gas humidifier.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2991-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 13, 2026
FDA report date
Sep 2, 2026
Quantity in commerce
20 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

All lot codes

View the original FDA recall notice

Why was this product recalled?

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

What actions did Medline Industries, LP take?

Medline issued an Urgent Medical Device notice it its consignees on 7/13/2026 via email and USPS. The notice explained the issue, potential risk, and provided actions and precautions for healthcare providers. If evidence of connector overheating or thermal damage is observed: " Assess the patient and ensure continued respiratory support. " Replace the affected circuit as soon as clinically appropriate. " Remove the affected circuit from service and retain it for investigation." For questions, contact the Recall Department at 866-359-1704 or [email protected]. Follow up notice sent on 7/27/2026 to clarify the item numbers. *** Updated customer communication was sent out on 8/5/26. Consignees were instructed to destroy affected inventory. The firm has begun distributing corrected product.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology