Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3004-2026

Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation Recall

Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation

FDA recall Class II

Ongoing

FDA lists this Class II recall for Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3004-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 22, 2026
FDA report date
Sep 2, 2026
Quantity in commerce
20 units
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation

Affected model & UDI checker

Search the exact identifiers published for Z-3004-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMerge PACS and Merge OrthoPACS, medical imaging diagnostic workstationCheck source notice00842000101239
FDA code information

Software version V9.5 - UDI (01)00842000101239(10)9.5.0(11)260409

View the original FDA recall notice

Why was this product recalled?

Software issue may cause measurements to be displayed inaccurately.

What actions did Merge Healthcare, Inc. take?

Merge Healthcare issued an Urgent Field Safety Notice to its consignees on 7/22/2026 via email with a follow-up telephone call. The notice explained the issue and how to identify it, risk to patient, and requested the following: "Immediate actions to reduce risk to patients Merge Healthcare recommends that customers contact Merge Healthcare Support to identify any affected ultrasound studies and the necessary next steps to be taken. Until the software update is installed, ultrasound measurements performed using the Merge PACS & OrthoPACS Viewer should be confirmed against the measurements acquired at the ultrasound modality." The required actions are as follows: "Merge Healthcare Support will contact all customers to: i) Confirm that you have read and understood this Urgent Field Safety Notice. ii) Assist you in identifying and reviewing any ultrasound studies with measurements that may have been affected. iii) Schedule the installation of the software update that eliminates the root cause of this issue. Please forward this notice to any individuals within the organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.