POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting & coagulation & accessories Recall
POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting & coagulation & accessories
FDA recall Class II
Ongoing
FDA lists this Class II recall for POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting & coagulation & accessories. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3035-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 18, 2026
- FDA report date
- Sep 2, 2026
- Quantity in commerce
- 335 units
- Distribution
- See key details
- FDA cause
- Labeling Change Control
What is being recalled?
POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting & coagulation & accessories
Affected model & UDI checker
Search the exact identifiers published for Z-3035-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| PS-0523CJDA | POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting & coagulation & accessories | Check source notice | Not specified |
FDA code information
Model Number: PS-0523CJDA. UDI-DI: 00821925044531. Lot Number: FR490902.
Why was this product recalled?
Potential labeling deficiency in which an outer shelf box is incorrectly labeled as a 37-cm device.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Olympus Corporation of the Americas take?
Olympus notified consignees on 08/18/2026 via "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were notified of the issue, hazard involved, and that the affected device could still be used. Consignees were instructed to review the recall notification and current inventory for the affected lot number, discard the outer box for any affected units on hand, and attach the recall notification letter to the inner device packaging. Consignees were also instructed to ensure all affected personnel are made aware, to ensure to check the sterile packaging and device labels before use to ensure you are using the correct length for your procedure, notify customers if products were further distributed and complete and return the provided response form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3035-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology