Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3037-2026

PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592 Recall

PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592

FDA recall Class II

Ongoing

FDA lists this Class II recall for PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3037-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 20, 2026
FDA report date
Sep 2, 2026
Quantity in commerce
4 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592

Affected model & UDI checker

Search the exact identifiers published for Z-3037-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
5700592PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592Check source noticeNot specified
Not specifiedPORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592Check source notice00198953056752
FDA code information

GTIN 00198953056752, Serial Numbers: SWT25510001TA, SWT26110010TA, SWT26070001TA, SWT26120008TA.

View the original FDA recall notice

Why was this product recalled?

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Ge Medical Systems Information Technologies, Inc. take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 7/20/2026 by letter delivered using traceable means. The notice explained the issue, potential risk, and requested the following: "Actions to be taken by Customer/User You can continue to use your Portrait Mobile Monitoring Solution with Portrait Core Services SW V1.0.5, V1.1 or V1.2. Please ensure appropriate alternative monitoring is established for patients before a software update is started. A User Manual addendum is included with this letter that includes specific instructions related to alternative monitoring. Please place a copy of the addendum in Appendix A with the User Manual for your Portrait Mobile Monitoring Solution with Portrait Core Services SW V1.0.5, V1.1 or V1.2 system in your facility. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology