Baxter Novum IQ Syringe Pump, Model Number 40800BAXUS Recall
Baxter Novum IQ Syringe Pump, Model Number 40800BAXUS
FDA recall Class II
Ongoing
FDA lists this Class II recall for Baxter Novum IQ Syringe Pump, Model Number 40800BAXUS. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3065-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 11, 2026
- FDA report date
- Sep 9, 2026
- Quantity in commerce
- 14601 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Baxter Novum IQ Syringe Pump, Model Number 40800BAXUS
Affected model & UDI checker
Search the exact identifiers published for Z-3065-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 40800BAXUS | Baxter Novum IQ Syringe Pump, Model Number 40800BAXUS | Check source notice | Not specified |
| Not specified | Baxter Novum IQ Syringe Pump, Model Number 40800BAXUS | Check source notice | 05413765852428 |
FDA code information
UDI-DI 05413765852428, all serial numbers
Why was this product recalled?
A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.
What actions did Baxter Healthcare Corporation take?
On August 11, 2026, firm began sending customers Urgent Medical Device Correction delivered via email and USPS first class mail. Customers were advised to use the same compatible syringe model (brand/size) for all syringes required within a Multi-Syringe VOT workflow. Customers were provided with additional instructions for situations where the syringe does not support the intended flow rate and the user receives a prompt to confirm continuation of the infusion at the maximum or minimum supported flow rate for that syringe after a syringe change.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3065-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.