Baxter Novum IQ Syringe Pump, Loaner, Model Number 40800BAXUSL Recall

Baxter Novum IQ Syringe Pump, Loaner, Model Number 40800BAXUSL

FDA recall Class II

Ongoing

FDA lists this Class II recall for Baxter Novum IQ Syringe Pump, Loaner, Model Number 40800BAXUSL. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3066-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 11, 2026
FDA report date
Sep 9, 2026
Quantity in commerce
4 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Baxter Novum IQ Syringe Pump, Loaner, Model Number 40800BAXUSL

Affected model & UDI checker

Search the exact identifiers published for Z-3066-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
40800BAXUSLBaxter Novum IQ Syringe Pump, Loaner, Model Number 40800BAXUSLCheck source noticeNot specified
Not specifiedBaxter Novum IQ Syringe Pump, Loaner, Model Number 40800BAXUSLCheck source notice00197886002300
FDA code information

UDI-DI 00197886002300, all serial numbers

View the original FDA recall notice

Why was this product recalled?

A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.

What actions did Baxter Healthcare Corporation take?

On August 11, 2026, firm began sending customers Urgent Medical Device Correction delivered via email and USPS first class mail. Customers were advised to use the same compatible syringe model (brand/size) for all syringes required within a Multi-Syringe VOT workflow. Customers were provided with additional instructions for situations where the syringe does not support the intended flow rate and the user receives a prompt to confirm continuation of the infusion at the maximum or minimum supported flow rate for that syringe after a syringe change.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology