LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model Number 7260, All Versions (Current 7.8) Recall
LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model Number 7260, All Versions (Current 7.8)
FDA recall Class II
Ongoing
FDA lists this Class II recall for LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model Number 7260, All Versions (Current 7.8). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3081-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 6, 2026
- FDA report date
- Sep 9, 2026
- Quantity in commerce
- N/A
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model Number 7260, All Versions (Current 7.8)
Affected model & UDI checker
Search the exact identifiers published for Z-3081-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 7260 | LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model Number 7260, All Versions (Current 7.8) | Check source notice | Not specified |
| Not specified | LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model Number 7260, All Versions (Current 7.8) | Check source notice | 00802526613838 |
FDA code information
UDI-DI: 00802526613838
Why was this product recalled?
There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did BOSTON SCIENTIFIC CORPORATION take?
An Urgent Medical Device Correction dated August 2026 was sent to customers. Boston Scientific is notifying you of the potential for a delay in AF Burden Alert notification affecting the LATITUDE" Clarity server software which supports the LUX-Dx" family of ICMs. This delayed AF Burden Alert notification behavior by LATITUDE Clarity may occur if the Alert for AF Burden e Threshold is configured to ">0 hours per day" setting (a non-default configuration). The ICM is performing as intended with continued AF detection, data storage, and transmission of device data. Boston Scientific has received one complaint related to this delayed notification behavior by the LATITUDE Clarity System. No patient injuries or deaths have been reported. A server software update is planned to correct this behavior. Configuration Options for Atrial Alert Notifications: Atrial arrhythmia alert notification preferences may vary, and AF configurations may be set for individual patients and/or group settings. Until updated server software is available, consider the following options: Change Threshold for AF Burden Alert Setting Enable AF Events Alert with Minimum Duration If Unchanged, Potential Delay in AF Burden Alert Notification " Boston Scientific does not recommend replacement of the implanted ICM. " Please share this notification with all HCPs within your organization who manage patients using the LUXDx ICM and LATITUDE Clarity System. Mitigation. The LATITUDE Clarity system is designed to detect atrial fibrillation through two independent mechanisms: AF Burden and AF Event Alerts. Until a software update to the LATITUDE Clarity server software is available, configuring AF Events or changing AF Burden Alerts settings can mitigate delays in full device interrogation and enable the LATITUDE Clarity system to display alert notification as intended. Additional Information. The Regulatory Authority of your country has been informed of this communication. Adverse events should be reported
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3081-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology