Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3084-2026

5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6 Recall

5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6

FDA recall Class II

Ongoing

FDA lists this Class II recall for 5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3084-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 10, 2026
FDA report date
Sep 9, 2026
Quantity in commerce
120 units
Distribution
See key details
FDA cause
Component change control

What is being recalled?

5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6

Affected model & UDI checker

Search the exact identifiers published for Z-3084-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specified5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6Check source notice00840861102556
FDA code information

Lots 25KR01227 25NR01257 UDI-DI 00840861102556

View the original FDA recall notice

Why was this product recalled?

Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.

What actions did Fresenius Medical Care Renal Therapies Group, LLC take?

On 8/10/2026, the firm began notifying consignees via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to 1. Identify any affected product on hand. 2. IMMEDIATELY segregate and discontinue use of affected lot. 3. Contact the call center at 1-888-943-4901 and reference Event 8514, between 8:00 AM 4:30 PM Eastern daily, for return of affected lots and replacement of your stock. 4, Complete and return the included reply form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology