Bravo CF capsule delivery device, 5-pack, Product Number FGS-0635, Bravo CF capsule delivery device, 1-pack, Product Number FGS-0636. Used to attach a pH monitoring capsule to the esophagus. Recall
Bravo CF capsule delivery device, 5-pack, Product Number FGS-0635, Bravo CF capsule delivery device, 1-pack, Product Number FGS-0636. Used to attach a pH monitoring capsule to the esophagus.
FDA recall Class I
Ongoing
FDA lists this Class I recall for Bravo CF capsule delivery device, 5-pack, Product Number FGS-0635, Bravo CF capsule delivery device, 1-pack, Product Number FGS-0636. Used to attach a pH monitoring capsule to the esophagus.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3090-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Aug 6, 2026
- FDA report date
- Sep 16, 2026
- Quantity in commerce
- 127,416
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Bravo CF capsule delivery device, 5-pack, Product Number FGS-0635, Bravo CF capsule delivery device, 1-pack, Product Number FGS-0636. Used to attach a pH monitoring capsule to the esophagus.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Product Number FGS-0635 UDI-DI (GTIN) 07290101369707 Product Number FGS-0636 UDI-DI (GTIN) 07290101369714 Expiration dates prior to and including October 20, 2027 (lot numbers from 64553F up to and including 68912F)
Why was this product recalled?
Capsule may fail to attach to the esophagus or detach from the delivery device. Risks include patient aspiration/inhalation, perforation of esophagus, obstruction of the airway, hemorrhage/blood loss/bleeding, laceration of the esophagus, a delay in diagnosis, and foreign bodies remaining in the patient.
What actions did Given Imaging Ltd. take?
On August 10, 2026, the firm notified consignees via "URGENT: MEDICAL DEVICE RECALL" letters. Please note this is a separate recall from that initiated in June 2025, which involved the same issue, but a different underlying cause. Customers were instructed to do the following: 1. Immediately identify and quarantine all unused and affected devices. 2. Return all product to Medtronic for replacement as described in the letter. 3. Complete and return the included customer response form to the firm via email: [email protected], even if you do not have unused affected inventory. 4. Notify all those who need to be aware within your organization, or those to whom affected product was transferred/distributed. 5. Maintain a copy of the recall letter for your records. If you have any questions regarding this communication, please contact your local Medtronic Representative or call 720-774-3454.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA classification recorded
FDA notice posted
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3090-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026