Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, LUNG TX CIRCULATOR Recall
Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, LUNG TX CIRCULATOR , Catalog Number: OVC3507231
FDA recall Class II
Ongoing
FDA lists this Class II recall for Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, LUNG TX CIRCULATOR. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3096-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 29, 2026
- FDA report date
- Sep 9, 2026
- Quantity in commerce
- 19 units
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, LUNG TX CIRCULATOR , Catalog Number: OVC3507231
Affected model & UDI checker
Search the exact identifiers published for Z-3096-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| OVC3507231 | Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, LUNG TX CIRCULATOR | Check source notice | Not specified |
FDA code information
1) OVC3507231, UDI-DI: 10198956232457(each), 50198956232455(case), Lot Number: NGKU0957
Why was this product recalled?
The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Cardinal Health 200, LLC take?
An URGENT: PRODUCT CORRECTION notification letter dated 7/29/26 was sent to customers. If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: [email protected] For component credit or replacement, please contact the following: " Hospitals Cardinal Customer Service at 866.551.0533 " Distributors Distributor Management at 800.635.6021 " Federal Government facilities Government Customer Service at 800.444.1166 If you are a Distributor and have further distributed affected kits please notify your customers of this recall. We sincerely apologize for any inconvenience this notice may have caused you and your staff. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3096-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology