Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3102-2026

Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYST Recall

Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYST , Catalog Number: PB41RHRM3; 2) KIT, ROBOTIC HYST , Catalog Number: PB41RHRM4; 3) KIT, C SECTION , Catalog Number: PLVMCS22G; 4) KIT, C SECTION MARTIN , Catalog Number: PL22CSMMI; 5) KIT, C SECTION MARTIN , Catalog Number: PL22CSMMI; 6) KIT, C SECTION OREGON HEALTH , Catalog Number: PL57CS60N; 7) C SECTION PACK , Catalog Number: SMAHBCSECB; 8) C SECTION PACK , Catalog Number: SMAHBCSEC3

FDA recall Class II

Ongoing

FDA lists this Class II recall for Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYST. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3102-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 29, 2026
FDA report date
Sep 9, 2026
Quantity in commerce
956 units
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYST , Catalog Number: PB41RHRM3; 2) KIT, ROBOTIC HYST , Catalog Number: PB41RHRM4; 3) KIT, C SECTION , Catalog Number: PLVMCS22G; 4) KIT, C SECTION MARTIN , Catalog Number: PL22CSMMI; 5) KIT, C SECTION MARTIN , Catalog Number: PL22CSMMI; 6) KIT, C SECTION OREGON HEALTH , Catalog Number: PL57CS60N; 7) C SECTION PACK , Catalog Number: SMAHBCSECB; 8) C SECTION PACK , Catalog Number: SMAHBCSEC3

Affected model & UDI checker

Search the exact identifiers published for Z-3102-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
PB41RHRM3Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYSTCheck source noticeNot specified
PB41RHRM4Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYSTCheck source noticeNot specified
PLVMCS22GPresource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYSTCheck source noticeNot specified
PL22CSMMIPresource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYSTCheck source noticeNot specified
PL57CS60NPresource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYSTCheck source noticeNot specified
SMAHBCSEC3Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYSTCheck source noticeNot specified
FDA code information

1) PB41RHRM3, UDI-DI: 10198956294219(each), 50198956294217(case), Lot Number: NGKU3049 ; 2) PB41RHRM3, UDI-DI: 10198956294219(each), 50198956294217(case), Lot Number: NGKU4043 ; 3) PB41RHRM4, UDI-DI: 10197106968413(each), 50197106968411(case), Lot Number: NGKU4716 ; 4) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0957 ; 5) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0856 ; 6) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0957 ; 7) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU2052 ; 8) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU2052 ; 9) PL22CSMMI, UDI-DI: 10195594952600(each), 50195594952608(case), Lot Number: NGKU0129 ; 10) PL22CSM13, UDI-DI: 10198956457553(each), 50198956457551(case), Lot Number: NGKU0129 ; 11) PL57CS60N, UDI-DI: 10197106889848(each), 50197106889846(case), Lot Number: NGKU2052 ; 12) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: NGKU4716 ; 13) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: NGKU4716 ; 14) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: 20251217 ; 15) SMAHBCSEC3, UDI-DI: 10198956267183(each), 50198956267181(case), Lot Number: NGKU2887

View the original FDA recall notice

Why was this product recalled?

The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Cardinal Health 200, LLC take?

An URGENT: PRODUCT CORRECTION notification letter dated 7/29/26 was sent to customers. If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: [email protected] For component credit or replacement, please contact the following: " Hospitals Cardinal Customer Service at 866.551.0533 " Distributors Distributor Management at 800.635.6021 " Federal Government facilities Government Customer Service at 800.444.1166 If you are a Distributor and have further distributed affected kits please notify your customers of this recall. We sincerely apologize for any inconvenience this notice may have caused you and your staff. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology