Halyard IV START KIT NON-STERILE REF: KRIV26-02 Recall
Halyard IV START KIT NON-STERILE REF: KRIV26-02
FDA recall Class I
Ongoing
FDA lists this Class I recall for Halyard IV START KIT NON-STERILE REF: KRIV26-02. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3108-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Jul 14, 2026
- FDA report date
- Sep 23, 2026
- Quantity in commerce
- 288 units
- Distribution
- See key details
- FDA cause
- Packaging process control
What is being recalled?
Halyard IV START KIT NON-STERILE REF: KRIV26-02
Affected model & UDI checker
Search the exact identifiers published for Z-3108-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| KRIV26-02 | Halyard IV START KIT NON-STERILE REF: KRIV26-02 | Check source notice | Not specified |
| Not specified | Halyard IV START KIT NON-STERILE REF: KRIV26-02 | Check source notice | 10809160460481 |
FDA code information
Lot Number: 1658437 UDI: 10809160460481
Why was this product recalled?
Convenience kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which are contaminated with Aspergillus penicillioides
What actions did AVID Medical, Inc. take?
On August 18, 2026, Avid Medical Inc. a subsidiary of Owens Minor issued an "Urgent Medical Device Recall Update" to affected consignees via email. The update was due to a revise recall strategy. Initial notification was sent on July 14, 2026. Avid ask consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product identified in TABLE 1. 2. Discontinue use of the BD ChloraPrep Clear - 1 mL and FREPP" Clear 1.5 mL Applicators. 3. FOR AVID KIT KRIV26-02, IV START KIT NON-STERILE, KIT LOT 1658437: Discard entire kit as it contains the contaminated component BD ChloraPrep" Clear - 1 mL, BD LOT 4032183. 4. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified of the actions to take and maintain records of effectiveness. 6. Complete the attached Recall Response Form and return by August 28, 2026. If you have any questions, contact Manager, Regulatory Affairs at 470-280-4277.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3108-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology