Medline Convenience Kits: 1) ACC010308A Recall
Medline Convenience Kits: 1) ACC010308A, Catalog Number: ACC010308A; 2) ACC010374A, Catalog Number: ACC010374A; 3) ACC010679, Catalog Number: ACC010679; 4) ACC010726A, Catalog Number: ACC010726A; 5) DYNDB3061, Catalog Number: DYNDB3061; 6) DYNDH1245, Catalog Number: DYNDH1245; 7) DYNDH1362, Catalog Number: DYNDH1362
FDA recall Class I
Ongoing
FDA lists this Class I recall for Medline Convenience Kits: 1) ACC010308A. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3124-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Aug 11, 2026
- FDA report date
- Sep 23, 2026
- Quantity in commerce
- 10407 kits
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Medline Convenience Kits: 1) ACC010308A, Catalog Number: ACC010308A; 2) ACC010374A, Catalog Number: ACC010374A; 3) ACC010679, Catalog Number: ACC010679; 4) ACC010726A, Catalog Number: ACC010726A; 5) DYNDB3061, Catalog Number: DYNDB3061; 6) DYNDH1245, Catalog Number: DYNDH1245; 7) DYNDH1362, Catalog Number: DYNDH1362
Affected model & UDI checker
Search the exact identifiers published for Z-3124-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| ACC010308A | Medline Convenience Kits: 1) ACC010308A | Check source notice | Not specified |
| ACC010374A | Medline Convenience Kits: 1) ACC010308A | Check source notice | Not specified |
| ACC010679 | Medline Convenience Kits: 1) ACC010308A | Check source notice | Not specified |
| ACC010726A | Medline Convenience Kits: 1) ACC010308A | Check source notice | Not specified |
| DYNDB3061 | Medline Convenience Kits: 1) ACC010308A | Check source notice | Not specified |
| DYNDH1245 | Medline Convenience Kits: 1) ACC010308A | Check source notice | Not specified |
| DYNDH1362 | Medline Convenience Kits: 1) ACC010308A | Check source notice | Not specified |
FDA code information
1) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 24HDA993; 2) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 24IDB285; 3) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 24JDA212; 4) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 24IDB849; 5) ACC010679, UDI-DI: 10195327396268(each), 40195327396269(case), Lot Number: 24HDA071; 6) ACC010726A, UDI-DI: 10198459043666(each), 40198459043667(case), Lot Number: 24GDB143; 7) DYNDB3061, UDI-DI: 10889942143733(each), 40889942143734(case), Lot Number: 24HBS752; 8) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24FMH385; 9) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24HBQ828
Why was this product recalled?
The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
What actions did Medline Industries, LP take?
An URGENT MEDICAL DEVICE RECALL notification letter dated 8/11/26 was sent to customers. This communication serves as an update to the Medline notification sent on June 12, 2026. Please be advised that the actions required for certain item(s) and lot(s) identified in the original notification have changed based on additional recall details received from BD (Becton, Dickinson and Company). REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-097-FGX3 Recall Code: 3.Your account will receive credit once the response form is submitted. 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3124-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology