Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3137-2026

Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N Recall

Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3137-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 23, 2026
FDA report date
Sep 23, 2026
Quantity in commerce
135 units
Distribution
See key details
FDA cause
Packaging process control

What is being recalled?

Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N

Affected model & UDI checker

Search the exact identifiers published for Z-3137-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMedline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008NCheck source notice10198459445941
Not specifiedMedline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008NCheck source notice40198459445942
FDA code information

UDI-DI: 10198459445941 (each), 40198459445942 (case); Lot Numbers: 25HLA462, 25HLA916, 25JLA847, 25LLA298, 26CLA372

View the original FDA recall notice

Why was this product recalled?

The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile barrier breach. Sterility of the product cannot be assured. If a compromised pouch seal is not identified prior to use and the cannula is used, there is a potential risk of infection or other complications associated with the use of a non-sterile device, including hemolysis, foreign body reaction, thromboembolism, and/or organ dysfunction.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medline Industries, LP take?

An URGENT MEDICAL DEVICE RECALL notification letter dated 7/23/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-121-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA classification recorded

  3. FDA notice posted

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology