FDA RECALL RECORD · Z-3142-2026

Philips PIC iX Essentials REF:867093 Recall

Philips PIC iX Essentials REF:867093

FDA recall Class II

Ongoing

FDA lists this Class II recall for Philips PIC iX Essentials REF:867093. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3142-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 31, 2026
FDA report date
Sep 23, 2026
Quantity in commerce
312 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Philips PIC iX Essentials REF:867093

Affected model & UDI checker

Search the exact identifiers published for Z-3142-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
867093Philips PIC iX Essentials REF:867093Check source noticeNot specified
Not specifiedPhilips PIC iX Essentials REF:867093Check source notice00884838122772
FDA code information

All PIC iX software versions 4.0-4.5/UDI: (01)00884838122772

View the original FDA recall notice

Why was this product recalled?

Due to software malfunction, patient information system may freeze and become unresponsive.

What actions did Philips North America take?

On July 31, 2026, Philips North America LLC issued an urgent medical device correction to affected consignees via USPS certified mail. Philips requested consignees to take the following actions: Users: 1. In cases where central monitoring has halted, please ensure that triaging of patients at bedside and a change in workflow occurs to more closely monitor patient status. Ensure audible alarm volume is on high in-patient rooms in case bedside devices disconnect and do not provide visual indicators that communication with the central. 2. Pass this notice to those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. 3. Place this Urgent notification with the documentation of the Philips Patient Information Center (PIC) iX and associate devices in a place where it is most likely to be seen and viewed. 4. Implement the Technical Solution established by Philips as soon as available within the defined timeframe. Distributors: 1. Send the attached URGENT Medical Device Correction to each customer to whom you have distributed any affected product(s) 2. Perform a good faith effort to reach each customer a minimum of three (3) times, and if possible, using multiple contact methods. 3. Implement the Technical Solution established by Philips as soon as available within the timeframe communicated by Philips, and confirm with Philips the implementation of the correction of all affected product(s). In case any product(s) cannot be corrected, inform Philips of the reason why the Technical Solution could not be implemented.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.