Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3162-2026

IV Sets with drip chambers Recall

IV Sets with drip chambers

FDA recall Class II

Ongoing

FDA lists this Class II recall for IV Sets with drip chambers. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3162-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 12, 2026
FDA report date
Sep 23, 2026
Quantity in commerce
86,113
Distribution
See key details
FDA cause
Process control

What is being recalled?

IV Sets with drip chambers

Affected model & UDI checker

Search the exact identifiers published for Z-3162-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedIV Sets with drip chambersCheck source notice20887787005831
FDA code information

All products manufactured on or before 7/4/2026. UDI-DI: 20887787005831

View the original FDA recall notice

Why was this product recalled?

Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Icu Medical, Inc. take?

On 5/12/2026, the firm posted a Information Regarding Reported Black Specks in Certain IV Tubing Set Drip Chambers from ICU Medical document on their website. On 5/18/2026, updated the notices were sent. On 7/28/2026, correction notices were sent to customers asking them to do the following: 1) Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used or further distributed to another location, please ensure that this communication is delivered there. 2) Complete and return the attached Customer Response Form to [email protected] 3) Continue using products as intended and per facility protocol. 4) Report any product quality concerns to ICU Medical Complaint Management. Firm contacts: [email protected] 1-866-261-8806. [email protected] 1-800-258-5361

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.