Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3173-2026

GM Helix Implant, REF: 109.984; Recall

GM Helix Implant, REF: 109.984;

FDA recall Class II

Ongoing

FDA lists this Class II recall for GM Helix Implant, REF: 109.984;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3173-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 12, 2026
FDA report date
Sep 23, 2026
Quantity in commerce
216
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

GM Helix Implant, REF: 109.984;

Affected model & UDI checker

Search the exact identifiers published for Z-3173-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
109.984GM Helix Implant, REF: 109.984;Check source noticeNot specified
FDA code information

REF: 109.984; UDI: (01)7899878024552(11)251122(17)301121(10)VZPF4; Lot: VZPF4;

View the original FDA recall notice

Why was this product recalled?

There is a discrepancy between the product identification on the packaging and the dimensions of the implant contained inside.

What actions did Straumann USA LLC take?

On August 12, 2026, Urgent Field Safety Action letters were sent to customers. Actions to be taken: 1. Identify and segregate any units in your stock with carton packaging labeled as Article 109.984 Helix GM Implant, 4.0 x 11.5 mm, L ot VZPF4 2. If the article is still in your inventory, then stop use/ distribution of the product immediately and quarantine / segregate physically. 3. Please return the affected items to the attention of Todd Shevlin at Straumann USA, LLC for replacement as indicated on the Customer Confirmation Form. 4. If the product has already been implanted, implant removal is not required, and no additional patient follow -up is required solely due to this recall 5. In all cases, complete and return the enclosed Customer Confirmation Form with the affected product using the return UPS label provided to: Straumann USA, LLC. If product is not being returned, the completed Customer Confirmation Form can be scanned and emailed to [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology