Blood Culture Collection Venipuncture Kit 20/Cs, REF: 77451; Blood Culture Kit 20/Cs, REF: 81834B; Blood Culture Collection Kit, REF: 81834C. All containing BD ChloraPrep Clear - 1 mL Applicators Recall
Blood Culture Collection Venipuncture Kit 20/Cs, REF: 77451; Blood Culture Kit 20/Cs, REF: 81834B; Blood Culture Collection Kit, REF: 81834C. All containing BD ChloraPrep Clear - 1 mL Applicators
FDA recall Class II
Ongoing
FDA lists this Class II recall for Blood Culture Collection Venipuncture Kit 20/Cs, REF: 77451; Blood Culture Kit 20/Cs, REF: 81834B; Blood Culture Collection Kit, REF: 81834C. All containing BD ChloraPrep Clear - 1 mL Applicators. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3181-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 23, 2026
- FDA report date
- Sep 23, 2026
- Quantity in commerce
- 6,420
- Distribution
- See key details
- FDA cause
- Material/Component Contamination
What is being recalled?
Blood Culture Collection Venipuncture Kit 20/Cs, REF: 77451; Blood Culture Kit 20/Cs, REF: 81834B; Blood Culture Collection Kit, REF: 81834C. All containing BD ChloraPrep Clear - 1 mL Applicators
Affected model & UDI checker
Search the exact identifiers published for Z-3181-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 77451 | Blood Culture Collection Venipuncture Kit 20/Cs, REF: 77451; Blood Culture Kit 20/Cs, REF: 81834B; Blood Culture Collection Kit, REF: 81834C. All containing BD ChloraPrep Clear - 1 mL Applicators | Check source notice | Not specified |
| 81834B | Blood Culture Collection Venipuncture Kit 20/Cs, REF: 77451; Blood Culture Kit 20/Cs, REF: 81834B; Blood Culture Collection Kit, REF: 81834C. All containing BD ChloraPrep Clear - 1 mL Applicators | Check source notice | Not specified |
| 81834C | Blood Culture Collection Venipuncture Kit 20/Cs, REF: 77451; Blood Culture Kit 20/Cs, REF: 81834B; Blood Culture Collection Kit, REF: 81834C. All containing BD ChloraPrep Clear - 1 mL Applicators | Check source notice | Not specified |
FDA code information
77451/20809160195069/343602(5/31/2027), 349005(10/31/2027), 374030(11/30/2027), 375434(12/31/2027), 377632(4/2/2028), 327194(8/31/2026), 375105(3/31/2028), 336074(3/21/2027); 81834B/20809160403324/326214(7/31/2027); 81834C/20809160466602/347551(8/12/2026), 372502(8/10/2026), 374615(10/21/2026), 376461(10/23/2026), 377088(11/6/2026), 373129(9/3/2026), 374033(9/3/2026)
Why was this product recalled?
Blood culture convenience kits contain a component subject to a drug recall that due to an an open or incomplete seal on the packaging of the applicator may compromise the component's sterile barrier and if used may lead to localized skin or soft tissue infection.
What actions did Medical Action Industries, Inc. 306 take?
On 7/23/2026, correction notices were emailed to customers who were asked to do the following: 1) Identify, segregate, and quarantine all impacted product. 2) Discontinue use of the BD ChloraPrep Clear - 1 mL Applicators 3) Add warning labels to affected MAI kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. 4) Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5) Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 6) Complete and return Recall Response Form via email to [email protected] Report any quality problems or adverse events associated with the products listed in this notification to [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3181-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology