REF:B12LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 12X100MM Recall
REF:B12LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 12X100MM
FDA recall Class II
Ongoing
FDA lists this Class II recall for REF:B12LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 12X100MM. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3185-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 19, 2026
- FDA report date
- Sep 23, 2026
- Quantity in commerce
- 2682 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
REF:B12LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 12X100MM
Affected model & UDI checker
Search the exact identifiers published for Z-3185-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| B12LT | REF:B12LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 12X100MM | Check source notice | Not specified |
FDA code information
UDI: (01)00885825020897/ Lots: 16029001 16032878 16038822 16041414 16044000 16048542 16052751 16054119 16062169 16062738 16066292 16074632 16087947 16087952 16090650 16098040 16107246 16111434 16115899 16115909 16120868 16130680 16130701 16135456 16136279 16138395 16140432 16140568 16142544 16142619 16159304 16160563 16172080 16176548 16186738 16186874 16187194 16187248 16187457 16197336 16201786 16205783 16205810 16205823 16206753 16220653 16220918 16221448 16221453
Why was this product recalled?
Reusable access port system may crack or seperate during use.
What actions did Stryker Sustainability Solutions take?
On August 19, 2026, Stryker Sustainability Solutions (SSS)issued a Urgent: Medical Device recall notice to affected consignees via sales representative and US Mail. SSS ask consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Immediately check your internal inventory to locate the affected product. Segregate the affected products, quarantine and discontinue use. 3. Complete the Business Reply Form or use the QR Code on Attachment A. Response is required. a.If you have indicated that product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to your facility for products returned upon receipt. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3185-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology