Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3193-2026

Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154; Recall

Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3193-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 20, 2026
FDA report date
Sep 23, 2026
Quantity in commerce
783
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;

Affected model & UDI checker

Search the exact identifiers published for Z-3193-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
5430-30151Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;Check source noticeNot specified
5430-30152Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;Check source noticeNot specified
5430-30154Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;Check source noticeNot specified
Not specifiedCaptus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;Check source notice00859942006096
Not specifiedCaptus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;Check source notice00859942006102
Not specifiedCaptus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;Check source notice00859942006423
FDA code information

Affected Units: Captus 4000: 1. Model 5430-30151; UDI-DI: 00859942006096; 2. Model 5430-30152; UDI-DI: 00859942006102; 3. Model 5430-30154; UDI-DI: 00859942006423; Devices shipped from April 2019 through February 2024, Serial Numbers 940548 through 941398, that have NOT already replaced the original arm with the redesigned arm.

View the original FDA recall notice

Why was this product recalled?

Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Mirion Technologies (Capintec), Inc. take?

On July 20, 2026 URGENT SAFETY NOTICE letters were emailed to customers. Actions to be taken: 1. Review this Alert and ensure that all affected personnel, including all operators of Captus 4000e Thyroid Uptake Systems, are aware of the contents. 2. Operate and inspect the device per the instructions provided. If your inspection shows that the collimator is not properly attached, please contact Capintec Technical Support at the phone number or email listed below. If you have any questions about this Inspection procedure of your Captus 4000 System, please contact Capintec Technical Support. If you notice any unusual change in the performance or functional response of the arm in your system (e.g. arm is making an unusual noise or arm does not move smoothly at any articulation joint), please contact Capintec Technical Support immediately. Capintec Technical Support: 1-800-631-3826 1-201-825-9500 [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology