SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyethylene) Recall
SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
FDA recall Class II
Ongoing
FDA lists this Class II recall for SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyethylene). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3203-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 30, 2026
- FDA report date
- Sep 23, 2026
- Quantity in commerce
- 12,156 US
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
3045-0026 No UDI Lots 12WAHA 12WAHB HOHE HOVX 3045-0028 No UDI Lots 13VAW 13VAWA 3045-0030 No UDI Lots BIYIE HNZY 3045-0032 No UDI Lots BIYIH HOYK
Why was this product recalled?
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did HOWMEDICA OSTEONICS CORP take?
Firm began notifying customers on July 30, 2026, via FedEx 2-day air. Customers were instructed to quarantine and discontinue use of the affected product. Affected product should be returned to the firm.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3203-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology