Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3208-2026

Convenience kits MINOR PACK DYNJ58421A SPRINGHILL MINOR PACK DYNJ40734F Recall

Convenience kits MINOR PACK DYNJ58421A SPRINGHILL MINOR PACK DYNJ40734F

FDA recall Class II

Ongoing

FDA lists this Class II recall for Convenience kits MINOR PACK DYNJ58421A SPRINGHILL MINOR PACK DYNJ40734F. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3208-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 31, 2026
FDA report date
Sep 23, 2026
Quantity in commerce
459
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Convenience kits MINOR PACK DYNJ58421A SPRINGHILL MINOR PACK DYNJ40734F

Affected model & UDI checker

Search the exact identifiers published for Z-3208-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedConvenience kits MINOR PACK DYNJ58421A SPRINGHILL MINOR PACK DYNJ40734FCheck source notice10193489970487
Not specifiedConvenience kits MINOR PACK DYNJ58421A SPRINGHILL MINOR PACK DYNJ40734FCheck source notice10198459162084
FDA code information

DYNJ58421A UDI-DI 10193489970487 Lot 25HBQ049 DYNJ40734F UDI-DI 10198459162084 Lots 25GBO299 25IBS607

View the original FDA recall notice

Why was this product recalled?

Kits may exhibit open seals in sterile pouch packaging. Open seal may compromise product sterility. If not detected prior to use, issue may lead to contamination of sterile field or patient infection if product is used.

What actions did Medline Industries, LP take?

Firm began notifying customers on July 31, 2026 via email and first-class mail. Customers were instructed to discard any affected sterile convenience kits.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology