Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Recall
Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3215-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 6, 2026
- FDA report date
- Sep 30, 2026
- Quantity in commerce
- 87
- Distribution
- See key details
- FDA cause
- Component design/selection
What is being recalled?
Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye
Affected model & UDI checker
Search the exact identifiers published for Z-3215-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| S3610 | Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module | Check source notice | Not specified |
| S2610 | Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module | Check source notice | Not specified |
| S2000 | Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module | Check source notice | Not specified |
| S1610 | Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module | Check source notice | Not specified |
| Not specified | Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module | Check source notice | 04059237000235 |
FDA code information
UDI-DI 04059237000235 Lot Code: Spec-07265, Spec-11121, Spec-11281, Spec-12096, Spec-13305, Spec-14403, Spec-15282, Spec-15587, Spec-20071, Spec-20348, Spec-20351, Spec-20591, Spec-20605, Spec-23473, Spec-23941, Spec-23950, Spec-23975, Spec-24507, Spec-24558, Spec-24582, Spec-24611
Why was this product recalled?
Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.
What actions did Heidelberg Engineering GmbH take?
Firm began notifying US customers on August 6, 2026 via email. Instruction to Customers: There is a possibility that the castor wheels of the Flex stand are not properly attached. The wheels may come loose when the Flex Module is being pulled or pushed during movement or transportation, affecting the stability of the stand. As a result, the Flex stand may tip over if not correctly handled or supported by the operator. Users shall schedule a service appointment with your Heidelberg Engineering service partner as soon as possible so that the castor fastenings can be inspected and, if necessary, corrected.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3215-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology