Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3215-2026

Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Recall

Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3215-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 6, 2026
FDA report date
Sep 30, 2026
Quantity in commerce
87
Distribution
See key details
FDA cause
Component design/selection

What is being recalled?

Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye

Affected model & UDI checker

Search the exact identifiers published for Z-3215-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
S3610Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex ModuleCheck source noticeNot specified
S2610Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex ModuleCheck source noticeNot specified
S2000Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex ModuleCheck source noticeNot specified
S1610Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex ModuleCheck source noticeNot specified
Not specifiedBrand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex ModuleCheck source notice04059237000235
FDA code information

UDI-DI 04059237000235 Lot Code: Spec-07265, Spec-11121, Spec-11281, Spec-12096, Spec-13305, Spec-14403, Spec-15282, Spec-15587, Spec-20071, Spec-20348, Spec-20351, Spec-20591, Spec-20605, Spec-23473, Spec-23941, Spec-23950, Spec-23975, Spec-24507, Spec-24558, Spec-24582, Spec-24611

View the original FDA recall notice

Why was this product recalled?

Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.

What actions did Heidelberg Engineering GmbH take?

Firm began notifying US customers on August 6, 2026 via email. Instruction to Customers: There is a possibility that the castor wheels of the Flex stand are not properly attached. The wheels may come loose when the Flex Module is being pulled or pushed during movement or transportation, affecting the stability of the stand. As a result, the Flex stand may tip over if not correctly handled or supported by the operator. Users shall schedule a service appointment with your Heidelberg Engineering service partner as soon as possible so that the castor fastenings can be inspected and, if necessary, corrected.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology