Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients Recall
Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients, Model Number: 1-200-0360 with Kit Number: 4-000-0724
FDA recall Class II
Ongoing
FDA lists this Class II recall for Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3220-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 21, 2026
- FDA report date
- Sep 30, 2026
- Quantity in commerce
- 33 units
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients, Model Number: 1-200-0360 with Kit Number: 4-000-0724
Affected model & UDI checker
Search the exact identifiers published for Z-3220-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 1-200-0360 | Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients | Check source notice | Not specified |
| Not specified | Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients | Check source notice | 00841709105838 |
| Not specified | Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients | Check source notice | 00841709105913 |
FDA code information
Model No: 1-200-0360; UDI-DI: 00841709105838; Kit No: 4-000-0724; UDI-DI: 00841709105913; Kit Serial Number: V2788703 V2788046 V2788016 V2788712 V2789302 V2789846 V2789830 V2789837 V2789808 V2789303 V2789304 V2811103 V2811112 V2862288 V2862275 V2869139 V2866577 V2867451 V2869599 V2868801 V2869154 V2868545 V2866984 V2869980 V2869969 V2868130 V2867433 V2868182 V2924525 V2924566 V2924559 V2924539 V2927525; Zone Serial Number: V2788322 V2787552 V2787555 V2788321 V2788905 V2789498 V2787554 V2789488 V2789499 V2788922 V2787553 V2807891 V2807893 V2837260 V2859050 V2868888 V2862992 V2866942 V2868893 V2868239 V2868840 V2868222 V2866940 V2868892 V2869397 V2867491 V2866963 V2868051 V2923639 V2922999 V2922998 V2923640 V2927201
Why was this product recalled?
A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Midmark Corporation take?
An Urgent Medical Device Recall notification letter dated 8/19/26 was sent to customers. Safety Instructions Until Midmark provides further instructions regarding this recall, use only the Step Deflation Method when obtaining blood pressure measurements. Do not use the Linear Deflation Method. Follow the instructions in Appendix A to verify that Step Deflation is selected before obtaining blood pressure measurements. Please complete, sign and return the Customer Response Form after completing the instructions described in Appendix A of this notification. The affected device(s) may continue to be used for blood pressure measurements provided this safety instruction is followed. Ensure all users of the affected device(s) are informed of this safety instruction. Product Correction Midmark will contact your facility with the next steps required to complete this recall. Additional instructions will be provided at that time. No immediate action is required to contact Midmark at this time. Contact Information Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact the Technical Service Team at 1.844.856.1230, option 2, or email [email protected]. Required Distributor Actions 1. Immediately send the attached customer-facing "Urgent Medical Device Recall - IQvitals Zone Linear Deflation Blood Pressure Performance" notification to every impacted end customer to whom you supplied an affected device. Do not wait for Midmark to contact the customer before forwarding the attached safety notification. 2. If you supplied an affected device to another distributor, send that distributor both this first-consignee communication and the attached customer notification. Instruct the downstream distributor to immediately forward only the customer notification to each impacted end customer and provide the final consignee information to you for submission to Midmark. 3. Immediately provide Midmark wi
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3220-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.