SeaSpine Mariner Deformity Product Name: Extended Polyaxial Head, Attached Gimbal, 12mm REF: MD1-300010-12 Recall

SeaSpine Mariner Deformity Product Name: Extended Polyaxial Head, Attached Gimbal, 12mm REF: MD1-300010-12

FDA recall Class II

Ongoing

FDA lists this Class II recall for SeaSpine Mariner Deformity Product Name: Extended Polyaxial Head, Attached Gimbal, 12mm REF: MD1-300010-12. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3226-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 19, 2026
FDA report date
Sep 30, 2026
Quantity in commerce
270 units
Distribution
See key details
FDA cause
Process design

What is being recalled?

SeaSpine Mariner Deformity Product Name: Extended Polyaxial Head, Attached Gimbal, 12mm REF: MD1-300010-12

Affected model & UDI checker

Search the exact identifiers published for Z-3226-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MD1-300010-12SeaSpine Mariner Deformity Product Name: Extended Polyaxial Head, Attached Gimbal, 12mm REF: MD1-300010-12Check source noticeNot specified
FDA code information

UDI-DI 10889981257248; Affected Lots: TE145060D, TE170356F, TE170556F, TE165793F, TE145059E, TE161254F, TE142282D.

View the original FDA recall notice

Why was this product recalled?

The proximal thread depth of the spinal device is out of design specification.

What actions did SeaSpine Orthopedics Corporation take?

On August 19, 2026, SeaSpine Orthopedics Corporation a subsidiary Orthofix issued a Urgent Medical Device Recall Notification via Email. SeaSpine ask consignees to take the following actions: Actions to be taken by the Customer/User: 1Discontinue use of affected lots 2Immediately identify and quarantine any affected product 3. If product is identified, Orthofix will facilitate the return 4.Complete and return the attached Acknowledgement and Receipt Form 5.This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.