Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3244-2026

Olympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Recall

Olympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Olympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3244-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 19, 2026
FDA report date
Sep 30, 2026
Quantity in commerce
404 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Olympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.

Affected model & UDI checker

Search the exact identifiers published for Z-3244-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
CD-B610LAOlympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector.Check source noticeNot specified
FDA code information

Model Number: CD-B610LA. UDI-DI: 00821925039469. Lot Numbers: All unexpired lots.

View the original FDA recall notice

Why was this product recalled?

Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Olympus Corporation of the Americas take?

Olumpus notified consignees on 08/19/2026 via "URGENT: MEDICAL DEVICE REMOVAL" letter. Consignees were informed of the reason for recall and hazard involved. Consignees were instructed to examine inventory for affected units, immediately cease use and quarantine them, arrange for their return, and confirm receipt of the recall notification through the portal. Consignees were also instructed to notify customers if product has been further distributed.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology