VentStar Resus Neo. Neonatal. 25 Units per Box. Part Number: MP00310 Disposable breathing circuit for use with a resuscitation module. Recall
VentStar Resus Neo. Neonatal. 25 Units per Box. Part Number: MP00310 Disposable breathing circuit for use with a resuscitation module.
FDA recall Class I
Ongoing
FDA lists this Class I recall for VentStar Resus Neo. Neonatal. 25 Units per Box. Part Number: MP00310 Disposable breathing circuit for use with a resuscitation module.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3252-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Sep 3, 2026
- FDA report date
- Oct 7, 2026
- Quantity in commerce
- 1,063,643 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
VentStar Resus Neo. Neonatal. 25 Units per Box. Part Number: MP00310 Disposable breathing circuit for use with a resuscitation module.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Part Number MP00310. UDIDI: 04048675422358. All lot numbers.
Why was this product recalled?
Potential cracks in the hose, which could lead to leakage or the hose falling off.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Draeger, Inc. take?
Draeger notified consignees on about 09/03/2026 via emailed Urgent Medical Device Recall letter. Consignees were informed of the issue of a cracked hose and that currently there are no alternative products from other manufacturers that have been tested and approved by Draeger for use with these devices, nor are there substitute products available within the Draeger portfolio in the U.S. Consignees were instructed to carefully inspect the hoses for damages prior to use according to the product s Instructions for Use (IFU), discard products that show cracks or have a detached hose, inform all potential users within the facility, and if affected products have been transferred or distributed, to notify those customers. Additionally, customers were requested to complete and return the provided Medical Device Recall Return Response Acknowledgement and Receipt Form. Draeger will replace the affected circuits free of charge after a correction is implemented based on the quantity indicated on the Medical Device Recall Return Response Acknowledgement and Receipt Form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA classification recorded
FDA notice posted
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3252-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology