nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R Recall
nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R
FDA recall Class II
Ongoing
FDA lists this Class II recall for nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3253-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 6, 2026
- FDA report date
- Sep 30, 2026
- Quantity in commerce
- 1,007
- Distribution
- See key details
- FDA cause
- Process design
What is being recalled?
nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
UDI-DI +B4388100EP1US02 All units with CO2 option SN beginning with 424, 125, 225, 325, 425, 126, 226, 326, 426, date range 2024-1st qt 2026
Why was this product recalled?
Complaint trend where "Checkline Error" alarm presents at device setup, rendering the CO2 monitoring function inoperable. May result in procedure delay.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Criticare Technologies, Inc. take?
Firm posted its initial notice on May 6, 2026. A follow-up letter was sent on August 31, 2026 with additional instructions. Customers were instructed to do the following: 1. Be aware of the Checkline Error alarm. If it appears on your CO2 section at startup, follow the steps provided in the letter. 2. Verify your backup plan. Confirm that your practice has access to an alternate CO2 monitoring source. 3. Retain the letter and note your unit s serial number for your records. 4. Contact the firm if you have any questions.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3253-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026