Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Recall
Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: VOLARA System Product Name: VOLARA, HC, NA A. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3257-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 3, 2026
- FDA report date
- Sep 30, 2026
- Quantity in commerce
- 2
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-3257-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| M08594A | Brand Name: VOLARA System Product Name: VOLARA, HC, NA A | Check source notice | Not specified |
| Not specified | Brand Name: VOLARA System Product Name: VOLARA, HC, NA A | Check source notice | 00887761981690 |
FDA code information
Lot Code: UDI: 00887761981690 Serial Numbers: W078SY3553 and Z309SY7853
Why was this product recalled?
Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.
What actions did Baxter Healthcare Corporation take?
Affected consignees were notified by phone on August 3, 2026. Customers were informed that as a precaution, Baxter would replace impacted VOLARA devices. A Baxter clinical trainer would contact the customers to schedule a home visit to exchange the device. Customers were asked to continue to use the VOLARA device as prescribed until it was exchanged. As of August 13, 2026, all potentially affected devices have been returned to Baxter.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3257-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology