Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Recall
Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Model Number: KRIV26-01
FDA recall Class II
Ongoing
FDA lists this Class II recall for Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3261-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 22, 2026
- FDA report date
- Sep 30, 2026
- Quantity in commerce
- 192 kits (4 cases)
- Distribution
- See key details
- FDA cause
- Packaging process control
What is being recalled?
Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Model Number: KRIV26-01
Affected model & UDI checker
Search the exact identifiers published for Z-3261-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| KRIV26-01 | Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit | Check source notice | Not specified |
FDA code information
Kit Model Number: KRIV26-01. UDI-DI: 10809160450109. Lot Number: 1642214. Expiration Date: 7/31/2027
Why was this product recalled?
Convenience kits containing BD ChloraPrep Clear - 1 mL Applicator, which may exhibit an open or incomplete seal on the packaging of the applicator.
What actions did AVID Medical, Inc. take?
Consignees were initially notified on about 07/22/2026 via "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter. They were informed that product may exhibit an open or incomplete seal on the packaging of the applicator. Consignees were instructed to quarantine impacted product, discontinue use of the BD ChloraPrep Clear - 1 mL Applicators within the kits, add a warning label to instruct users to discard the impacted component, and notify customers if the kits had been further distributed. Consignees were notified again on about 08/20/2026 with an updated "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter. Consignees were informed that certain lots of the BD ChloraPrep Applicators are contaminated with Aspergillus penicillioides. Instructions included to discard impacted kits containing affected applicators and notify customers if the kits had been further distributed. For those unaffected by the contamination, but are included in the action regarding sterility concerns, they were instructed to quarantine impacted product, discontinue use of the BD ChloraPrep Clear - 1 mL Applicators within the kits, add a warning label to instruct users to discard the impacted component, and notify customers if the kits had been further distributed.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3261-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology