FDA RECALL RECORD · Z-3279-2026

Computed Tomography X-ray System, Model: uCT 760 with software Recall

Computed Tomography X-ray System, Model: uCT 760 with software

FDA recall Class II

Ongoing

FDA lists this Class II recall for Computed Tomography X-ray System, Model: uCT 760 with software. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3279-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 11, 2026
FDA report date
Oct 7, 2026
Quantity in commerce
1
Distribution
See key details
FDA cause
Software design

What is being recalled?

Computed Tomography X-ray System, Model: uCT 760 with software

Affected model & UDI checker

Search the exact identifiers published for Z-3279-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedComputed Tomography X-ray System, Model: uCT 760 with softwareCheck source notice06971576831043
FDA code information

UDI-DI: 06971576831043. Serial Number/Software: 600271/R003.8.4.1810621-Re3300SP1H1_P3

View the original FDA recall notice

Why was this product recalled?

Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Shanghai United Imaging Healthcare Co., Ltd. take?

On 8/11/2026, correction notices were emailed to customers informing them that software will be released to resolve the software problem. Customers were asked to do the following: When using the Study Split function, follow the requirements below: 1) Temporarily disable the "Study Split" function. Navigate to the system configuration interface and uncheck the "Auto Study Split" option. 2) If the "Study Split" function is clinically necessary, open the system configuration interface and select "Split on PACS" instead of "Split on Local". 3) When "Auto study split" is checked, ensure that PACS (Picture Archiving and Communication System) is configured on site and auto-archive is enabled. 4) If images are found to be missing on the PA interface, immediately verify whether the corresponding image data exists in the PACS. a) If images exist in PACS: They can be retrieved from PACS for review. b) If images do not exist in PACS: Contact the firm's service engineer for assistance. For further questions, contact firm service engineer or call firm Customer Contact Center at (855) 221-1552 or 832-699-2799

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.