Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3291-2026

Brand Name: Opportunity Pro Product Name: Opportunity Pro Recall

Brand Name: Opportunity Pro Product Name: Opportunity Pro Model/Catalog Number: 72-50800, 72-50801,72-50802,72-50803,72-50804 Product Description: Blue Nitrile patient examination glove

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Opportunity Pro Product Name: Opportunity Pro. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3291-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 14, 2026
FDA report date
Oct 7, 2026
Quantity in commerce
17,481,000
Distribution
See key details
FDA cause
Process control

What is being recalled?

Brand Name: Opportunity Pro Product Name: Opportunity Pro Model/Catalog Number: 72-50800, 72-50801,72-50802,72-50803,72-50804 Product Description: Blue Nitrile patient examination glove

Affected model & UDI checker

Search the exact identifiers published for Z-3291-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
72-50800Brand Name: Opportunity Pro Product Name: Opportunity ProCheck source noticeNot specified
72-50801Brand Name: Opportunity Pro Product Name: Opportunity ProCheck source noticeNot specified
72-50802Brand Name: Opportunity Pro Product Name: Opportunity ProCheck source noticeNot specified
72-50803Brand Name: Opportunity Pro Product Name: Opportunity ProCheck source noticeNot specified
72-50804Brand Name: Opportunity Pro Product Name: Opportunity ProCheck source noticeNot specified
FDA code information

Item No. 72-RM50800 UDI-DI 818634025487 Lots AO01519 Item No. 72-RM50801 UDI-DI 818634025401 Lots AO01518 AO01564 Item No. 72-RM50802 UDI-DI 818634025418 Lots AO01517 AO01531 AO01530 AO01568 Item No. 72-RM50803 UDI-DI 818634025425 Lots AO01529 AO01567 AO01593 AO01594 Item No. 72-RM50804 UDI-DI 818634025432 Lots AO01222 AO01528 AO01566 Alternate lots AN250310948, AN250310947, AN250310945, AN250311327

View the original FDA recall notice

Why was this product recalled?

Gloves may exhibit defects such as tears, missing fingers, and debris. Use of defective gloves may lead to healthcare personnel and patient exposure to infectious pathogens leading to infection or cross-contamination, or exposure to chemical agents leading to irritation or systemic toxicities.

What actions did American Nitrile Operations Llc take?

Firm began notifying its VA customers on August 15, 2026 via email. Instruction to Customers: 1. Immediately cease all use of the affected products. 2. Notify all users throughout your facility of the product recall. 3. Examine your inventory and immediately quarantine the affected lots to avoid accidental use 4 Contact distributor for replacement.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.