Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ROBOTIC PACK - 242561 Model No. MHRB88AO Recall
Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ROBOTIC PACK - 242561 Model No. MHRB88AO
FDA recall Class II
Ongoing
FDA lists this Class II recall for Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ROBOTIC PACK - 242561 Model No. MHRB88AO. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3292-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 10, 2026
- FDA report date
- Oct 7, 2026
- Quantity in commerce
- 1,804 kits total
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ROBOTIC PACK - 242561 Model No. MHRB88AO
Affected model & UDI checker
Search the exact identifiers published for Z-3292-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MHRB88AO | Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ROBOTIC PACK - 242561 Model No. MHRB88AO | Check source notice | Not specified |
| Not specified | Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ROBOTIC PACK - 242561 Model No. MHRB88AO | Check source notice | 191072236531 |
FDA code information
UDI-DI 191072236531 Lots 42-8513411 42-8580011 42-8663211 42-8775711 42-8801311 42-8840611 42-8865911
Why was this product recalled?
Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a contaminated wipe may lead to localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare due to disposal of contaminated wipes may increase risk of localized infection or catheter-associated urinary tract infection (CAUTI).
What actions did AMERICAN CONTRACT SYSTEMS, INC. take?
Firm began notifying affected distributors via email on August 10, 2026. Distributors were instructed to notify end-users. Customers were instructed to identify, segregate, and quarantine all impacted products. Customers are to add recall labels to affected kits to ensure the impacted component is discarded prior to use. An amended customer letter was sent out on August 28, 2026, which added the following instruction: Users should source alternative cleansing wipes for the impacted product lots and ensure they continue to follow their facility protocols for providing perineal care for patients with urinary catheters.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3292-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology