Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3300-2026

Olympus M3 Gold Autoclavable Lateral Telescope 70. Recall

Olympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Olympus M3 Gold Autoclavable Lateral Telescope 70.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3300-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 24, 2026
FDA report date
Oct 7, 2026
Quantity in commerce
50 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Olympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.

Affected model & UDI checker

Search the exact identifiers published for Z-3300-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
M3-70AOlympus M3 Gold Autoclavable Lateral Telescope 70.Check source noticeNot specified
FDA code information

Model Number: M3-70A. UDI-DI: 00821925005686. All Serial Numbers.

View the original FDA recall notice

Why was this product recalled?

There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.

What actions did Olympus Corporation of the Americas take?

Olympus notified the single Japanese distributor on about 08/24/2026 via emailed "URGENT: FIELD SAFETY NOTICE." Consignees and end users are being instructed to store a copy of the provided correct sterilization parameters with current package insert, use only the corrected sterilization parameters, complete and return the provided reply form, and if units have been further distributed, notify those customers.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology