FDA RECALL RECORD · Z-3317-2026

Upgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13; Recall

Upgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Upgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3317-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 28, 2026
FDA report date
Oct 7, 2026
Quantity in commerce
30
Distribution
See key details
FDA cause
Software design

What is being recalled?

Upgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;

Affected model & UDI checker

Search the exact identifiers published for Z-3317-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedUpgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;Check source notice00884838104402
FDA code information

Model number (REF): 782130; UDI-DI: 00884838104402; Serial Numbers: 42002, 42030, 42037, 42066, 42104, 42151, 42163, 42164, 42211, 42219, 42248, 42376, 45001, 45007, 45043, 45044, 45052, 45179, 45236, 45263, 45552, 71060, 71262, 71279, 71406, 71532, 71565, 71580, 71616, 71907;

View the original FDA recall notice

Why was this product recalled?

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

What actions did Philips North America take?

On August 28, 2026, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. To prevent the issue from occurring while Philips is developing the software solution, use the Classic DICOM transfer option for exporting DTI scans. 2. You may continue to use your system(s) in accordance with the intended use. 3. Circulate this notice to all users of this device so that they are aware of the issue and associated hazard/harm. 4. Please retain this Urgent Medical Device Correction letter with your system(s) until the software upgrade is installed; ensure the notice is in a place likely to be seen/viewed.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology