Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Recall
Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3330-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 16, 2026
- FDA report date
- Oct 7, 2026
- Quantity in commerce
- 177
- Distribution
- See key details
- FDA cause
- Software Design Change
What is being recalled?
Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-3330-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS | Check source notice | 00389152470641 |
FDA code information
Lot Code: UDI-DI: 00389152470641 Software Version: 2.13 (iOS only)
Why was this product recalled?
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.
What actions did Tandem Diabetes Care, Inc. take?
On 07/16/2026, an "URGENT MEDICAL DEVICE CORRECTION" communication was sent via certified mail to customers informing them that an unexpected behavior was identified in the t:slim iOS mobile application version 2.13 made available through TestFlight. The issue involves users unable to pair a compatible FreeStyle Libre 3 Plus sensor after updating to version 2.13. If this occurs, the application will incorrectly display an "Incompatible Sensor" message and prevent sensor use. Customer are advised to: 1. Discontinue use of iOS t:slim mobile application version 2.13 2. Uninstall version 2.13 from their mobile device. 3. Return to the current production version of the t:slim mobile application (version 2.12.1) available on the Apple App Store. 4. Scan the compatible FreeStyle Libre 3 Plus sensor with version 2.12.1 when starting their next sensor session. For Questions - contact Tandum Customer Support at 1-877-801-6901 or email [email protected] (24/7/365).
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3330-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.